Calibration curve

Calibration curve

In analytical chemistry, a calibration curve is a general method for determining the concentration of a substance in an unknown sample by comparing the unknown to a set of standard samples of known concentration.

The calibration curve is a plot of how the instrumental response, the so-called “analytical signal,” changes with changing concentration of analyte (substance to be measured). The operator will create a series of standards across a range of concentrations near the expected unknown concentration. One must take care that these concentrations are in the working range of the technique (instrumentation) they are using. Analyzing each of these standards using the chosen technique will produce a series of readings. For most analyses a plot of response vs. concentration will create a linear relationship. The operator can measure the response of the unknown and, using the calibration curve, can "interpolate" to find the concentration of analyte.

How to create a calibration curve

To find the best-fit straight line we use linear regression analysis. From the equation y = mx + c you substitute in the y value (response) and solve for x.

Many different variables can be used as the analytical signal, in Fig. 1 analysis was of Chromium(III) by chemiluminescence. The detector (PMT) detects light, voltage increases with intensity of light, this creates a peak and the peak area is the analytical signal.

For most analytical techniques, the ultimate goal is to obtain a calibration curve; and there are a number of advantages to this method:

The calibration curve not only gives you your answer it also gives you an idea about how good that answer is. Provided you are operating in the linear response range the plot should be a straight line, deviations from this straight line give a good indication about the precision of the result. That is to say, if the data points are spread out, there is less certainty in the result. [For a detailed discussion on how to find a numerical error from a calibration curve see any general analytical chemistry textbook. For example Principles of Instrumental Analysis by D.A. Skoog.]

The calibration curve provides you with an empirical relationship, as opposed to a theoretical one. Instrumental response is usually highly dependent on the condition of the analyte, solvents used and impurities it may contain; it could also be affected by external factors such as pressure and temperature. Many theoretical relationships, such as fluorescence, require the determination of an instrumental constant anyway; in which case a calibration curve is the only way to do the determination. By making standards as similar as possible to the unknown and creating the calibration curve, what is created is a custom relationship, which takes into account all the above factors and many others for the specific conditions of the experiment.

The disadvantages are that a set of standards needs to be made, for which you must have a source of analyte material of known composition. This is a time-consuming process, may use significant quantities of expensive chemicals, and results in an increased amount of waste for disposal. Also, you need to make a rough estimate of the composition of your unknown sample, in order to enable interpolation the concentrations of the standards should fall above and below the expected unknown concentration. In short, there is a lot of information about the analysis in a calibration curve and it is a simple way to account for all the influencing factors that could change the analytical signal.

Applications

*Analysis of concentration
*Verifying the proper functioning of an analytical instrument or a sensor device such as an ion selective electrode
*Determining the basic effects of a control treatment (such as a dose-survival curve in clonogenic assay)

Notes

Bibliography

*cite book |author=Harris, Daniel Charles |title=Quantitative chemical analysis |publisher=W.H. Freeman |location=San Francisco |year=2003 |pages= |isbn=0-7167-4464-3 |oclc= |doi=
*cite book |author=Crouch, Stanley; Skoog, Douglas A. |title=Principles of Instrumental Analysis |publisher=Brooks Cole |location=Pacific Grove |year= |pages= |isbn=0-495-01201-7 |oclc= |doi=
*cite journal |author=Lavagnini I, Magno F |title=A statistical overview on univariate calibration, inverse regression, and detection limits: Application to gas chromatography/mass spectrometry technique |journal=Mass spectrometry reviews |volume=26 |issue=1 |pages=1–18 |year=2007 |pmid=16788893 |doi=10.1002/mas.20100


Wikimedia Foundation. 2010.

Игры ⚽ Поможем решить контрольную работу

Look at other dictionaries:

  • calibration curve — kalibravimo kreivė statusas T sritis Standartizacija ir metrologija apibrėžtis Matuoklio padalų gradavimo, jo paklaidų įvertinimo ar matavimo rezultatų perskaičiavimo kitais matavimo vienetais kreivė. atitikmenys: angl. calibration curve; gaging… …   Penkiakalbis aiškinamasis metrologijos terminų žodynas

  • calibration curve — kalibravimo kreivė statusas T sritis fizika atitikmenys: angl. calibration curve; gauging curve vok. Eichkurve, f rus. калибровочная кривая, f pranc. courbe de calibrage, f; courbe de calibration, f …   Fizikos terminų žodynas

  • calibration curve — gradavimo kreivė statusas T sritis automatika atitikmenys: angl. calibration curve; graduation curve vok. Eichkurve, f rus. градуировочная кривая, f pranc. courbe de graduation, f …   Automatikos terminų žodynas

  • calibration curve — kalibravimo kreivė statusas T sritis chemija apibrėžtis Matuoklio padalų verčių nustatymo, paklaidų įvertinimo ar matavimo rezultatų perskaičiavimo į kitus vienetus kreivė. atitikmenys: angl. calibration curve rus. калибровочная кривая …   Chemijos terminų aiškinamasis žodynas

  • Calibration — is the process of establishing the relationship between a measuring device and the units of measure. This is done by comparing a device or the output of an instrument to a standard having known measurement characteristics. For example the length… …   Wikipedia

  • curve — 1. A nonangular continuous bend or line. 2. A chart or graphic representation, by means of a continuous line connecting individual observations, of the course of a physiologic activity, of the number of cases of a disease in a given period, or of …   Medical dictionary

  • courbe de calibration — kalibravimo kreivė statusas T sritis Standartizacija ir metrologija apibrėžtis Matuoklio padalų gradavimo, jo paklaidų įvertinimo ar matavimo rezultatų perskaičiavimo kitais matavimo vienetais kreivė. atitikmenys: angl. calibration curve; gaging… …   Penkiakalbis aiškinamasis metrologijos terminų žodynas

  • gaging curve — kalibravimo kreivė statusas T sritis Standartizacija ir metrologija apibrėžtis Matuoklio padalų gradavimo, jo paklaidų įvertinimo ar matavimo rezultatų perskaičiavimo kitais matavimo vienetais kreivė. atitikmenys: angl. calibration curve; gaging… …   Penkiakalbis aiškinamasis metrologijos terminų žodynas

  • gauging curve — kalibravimo kreivė statusas T sritis Standartizacija ir metrologija apibrėžtis Matuoklio padalų gradavimo, jo paklaidų įvertinimo ar matavimo rezultatų perskaičiavimo kitais matavimo vienetais kreivė. atitikmenys: angl. calibration curve; gaging… …   Penkiakalbis aiškinamasis metrologijos terminų žodynas

  • courbe de calibration — kalibravimo kreivė statusas T sritis fizika atitikmenys: angl. calibration curve; gauging curve vok. Eichkurve, f rus. калибровочная кривая, f pranc. courbe de calibrage, f; courbe de calibration, f …   Fizikos terminų žodynas

Share the article and excerpts

Direct link
Do a right-click on the link above
and select “Copy Link”