Dosimetry

Dosimetry

Radiation dosimetry is the measurement and calculation of the absorbed dose in matter and tissue resulting from the exposure to indirect and direct ionizing radiation. It is a scientific subspecialty in the fields of health physics and medical physics that is focused on the calculation of internal and external doses from ionizing radiation.

Dose is reported in gray (Gy) for matter or sieverts (Sv) for biological tissue, where 1 Gy or 1 Sv is equal to 1 joule per kilogram. Non-SI units are still prevalent as well, where dose is often reported in rads and dose equivalent in rems. By definition, 1 Gy = 100 rad and 1 Sv = 100 rem.

In the United States, radiation dosimeters measure and display dose in rems rather than sieverts. One rem = 1 centisievert = 1/100 sievert. Popular emergency-level personal radiation dosimeters are DOSE-GARD and EPD N2, made by General Atomics Inc. and Thermoscientific, respectively.

Contents

Radiation effects on living tissue

The distinction between absorbed dose (Gy) and dose equivalent (Sv) is based upon the biological effects of the weighting factor (denoted wr) and tissue/organ weighting factor (WT) have been established, which compare the relative biological effects of various types of radiation and the susceptibility of different organs.

Organ dose weighting factors

By definition, the weighting factor for the whole body is 1, such that 1 Gy of radiation delivered to the whole body (i.e. an evenly distributed 1 joule of energy deposited per kilogram of body) is equal to one sievert (for photons with a radiation weighting factor of 1, see below). Therefore, the weighting factors for each organ must sum to 1 as the unit gray is defined per kilogram and is therefore a local effect. As the table below shows, 1 gray (photons) delivered to the gonads is equivalent to 0.08 Sv to the whole body—in this case, the actual energy deposited to the gonads, being small, would also be small.

Organ or tissue WT
Gonads 0.25
Breasts 0.15
Red Bone Marrow 0.12
Lung 0.12
Thyroid 0.03
Bone surfaces 0.03
Remainder 0.30
Whole body 1.0

Radiation weighting factors

By definition, x-rays and gamma rays have a weighting factor of unity, such that 1 Gy = 1 Sv (for whole-body irradiation). Values of wr are as high as 20 for alpha particles and neutrons, i.e. for the same absorbed dose in Gy, alpha particles are 20 times as biologically potent as X or gamma rays.

Dose versus activity

Radiation dose refers to the amount of energy deposited in matter and/or biological effects of radiation, and should not be confused with the unit of radioactive activity (becquerel, Bq). Exposure to a radioactive source will give a dose which is dependent on the activity, time of exposure, energy of the radiation emitted, distance from the source and shielding. The equivalent dose is then dependent upon the weighting factors above. Dose is a measure of deposited dose, and therefore can never go down—removal of a radioactive source can only reduce the rate of increase of absorbed dose, never the total absorbed dose.

The worldwide average background dose for a human being is about 3.5 mSv per year [1], mostly from cosmic radiation and natural isotopes in the earth. The largest single source of radiation exposure to the general public is naturally-occurring radon gas, which comprises approximately 55% of the annual background dose. It is estimated that radon is responsible for 10% of lung cancers in the United States.

Measuring dose

There are several ways of measuring doses from ionizing radiation. Workers who come in contact with radioactive substances or may be exposed to radiation routinely carry personal dosimeters. In the United States, these dosimeters usually contain materials that can be used in thermoluminescent dosimetry (TLD) or optically stimulated luminescence (OSL). Outside the United States, the most widely-used type of personal dosimeter is the film badge dosimeter, which uses photographic emulsions that are sensitive to ionizing radiation. The equipment used in radiotherapy (linear particle accelerator in external beam therapy) is routinely calibrated using ionization chambers or the new and more accurate diode technology.

Dose standards

Because the human body is approximately 70% water and has an overall density close to 1 g/cm3, dose measurement is usually calculated and calibrated as dose to water.

National standards laboratories such as the NPL provide calibration factors for ionization chambers and other measurement devices to convert from the instrument's readout to absorbed dose. The standards laboratories operate a Primary Standard, which is normally calibrated by absolute calorimetry, the warming of substances when they absorb energy. A user sends their Secondary Standard to the laboratory, where it is exposed to a known amount of radiation (derived from the Primary Standard) and a factor is issued to convert the instrument's reading to that dose. The user may then use their Secondary Standard to derive calibration factors for other instruments they use, which then become tertiary standards, or field instruments.

The NPL in the UK operates a graphite-calorimeter for absolute photon dosimetry. Graphite is used instead of water as its specific heat capacity is one-sixth that of water and therefore the temperature rises in graphite are 6 times more than the equivalent in water and measurements are more accurate. Significant problems exist in insulating the graphite from the laboratory in order to measure the tiny temperature changes. A lethal dose of radiation to a human is approximately 10–20 Gy. This is 10-20 joules per kilogram. A 1 cm3 piece of graphite weighing 2 grams would therefore absorb around 20–40 mJ. With a specific heat capacity of around 700 J·kg−1·K−1, this equates to a temperature rise of just 20 mK.

Medical dosimetry

Medical dosimetry is the calculation of absorbed dose and optimization of dose delivery in radiation therapy. It is often performed by a professional medical dosimetrist with specialized training in the field. In order to plan the delivery of radiation therapy, the radiation produced by the sources is usually characterized with percentage depth dose curves and dose profiles measured by medical physicists.

Dosimetry in NMR

Improperly said NMR dosimeter, the device shown in picture is a sampler for monitoring the static magnetic field in Nuclear Magnetic Resonance. This device is an electrical SMD device that store data for post processing analysis and evaluation of risk exposed operators.

Dosimeter TALETE for RMN

See also

External links


Wikimedia Foundation. 2010.

Игры ⚽ Нужно решить контрольную?

Look at other dictionaries:

  • dosimetry — do*sim e*try, n. [NL. dosis dose + metry.] (Med.) Measurement of doses; specif., a system of therapeutics which uses but few remedies, mostly alkaloids, and gives them in doses fixed by certain rules. {Do si*met ric}, a. {Do*sim e*trist}, n.… …   The Collaborative International Dictionary of English

  • Dosimetry — См. Дозиметрия Термины атомной энергетики. Концерн Росэнергоатом, 2010 …   Термины атомной энергетики

  • dosimetry — dozimetrija statusas T sritis Standartizacija ir metrologija apibrėžtis Dozių matavimas ar skaičiavimas. atitikmenys: angl. dosage measurement; dosimetry vok. Dosimetrie, f; Dosismessung, f rus. дозиметрия, f pranc. dosimétrie, f …   Penkiakalbis aiškinamasis metrologijos terminų žodynas

  • dosimetry — dozimetrija statusas T sritis Standartizacija ir metrologija apibrėžtis Jonizuojančiosios spinduliuotės ir jos sąveikos su gyvąja gamta charakteristikų matavimai ir tyrimai. Matuoja spinduliuotės dozes, radioaktyviųjų preparatų aktyvumą, kuria… …   Penkiakalbis aiškinamasis metrologijos terminų žodynas

  • dosimetry — dozimetrija statusas T sritis Standartizacija ir metrologija apibrėžtis Mokslas, tiriantis jonizuojančiosios spinduliuotės sąveiką su gyvąja ir negyvąja gamta, tą sąveiką apibūdinančių dydžių matavimo metodus ir matuoklius. atitikmenys: angl.… …   Penkiakalbis aiškinamasis metrologijos terminų žodynas

  • dosimetry — dozimetrija statusas T sritis chemija apibrėžtis Radioaktyviųjų medžiagų aktyvumo matavimo metodai. atitikmenys: angl. dosimetry rus. дозиметрия …   Chemijos terminų aiškinamasis žodynas

  • dosimetry — dozimetrija statusas T sritis fizika atitikmenys: angl. dosage measurement; dosimetry vok. Dosimetrie, f; Dosismeßung, f rus. дозиметрия, f; измерение дозы, n pranc. dosimétrie, f; mesure de dose, f …   Fizikos terminų žodynas

  • dosimetry — dozimetrija statusas T sritis apsauga nuo naikinimo priemonių apibrėžtis Apšvitos dozių matavimo metodai. Šiuo terminu apibūdinami dozimetriniai prietaisai ir jų naudojimo metodika. atitikmenys: angl. dosimetry rus. дозиметрия …   Apsaugos nuo naikinimo priemonių enciklopedinis žodynas

  • dosimetry — dozimetrija statusas T sritis apsauga nuo naikinimo priemonių apibrėžtis Branduolio fizikos šaka, tirianti fizikinius dydžius, kurie apibūdina jonizuojančiosios spinduliuotės sąveiką su gyvąja ir negyvąja gamta, tų dydžių matavimo metodus ir… …   Apsaugos nuo naikinimo priemonių enciklopedinis žodynas

  • dosimetry — dozimetrija statusas T sritis ekologija ir aplinkotyra apibrėžtis Jonizuojančiosios spinduliuotės ir jos sąveikos su gyvąja gamta charakteristikų matavimas ir tyrimas. atitikmenys: angl. dosage; dosimetry; measurement; monitoring vok. Dosimetrie …   Ekologijos terminų aiškinamasis žodynas

Share the article and excerpts

Direct link
Do a right-click on the link above
and select “Copy Link”