Contrast-enhanced ultrasound

Contrast-enhanced ultrasound

Contrast-enhanced ultrasound (CEUS) is the application of ultrasound contrast medium to traditional medical sonography. Ultrasound contrast agents rely on the different ways in which sound waves are reflected from interfaces between substances. This may be the surface of a small air bubble or a more complex structure. Commercially available contrast media are gas-filled microbubbles that are administered intravenously to the systemic circulation. Microbubbles have a high degree of echogenicity, which is the ability of an object to reflect the ultrasound waves. The echogenicity difference between the gas in the microbubbles and the soft tissue surroundings of the body is immense. Thus, ultrasonic imaging using microbubble contrast agents enhances the ultrasound backscatter, or reflection of the ultrasound waves, to produce a unique sonogram with increased contrast due to the high echogenicity difference. Contrast-enhanced ultrasound can be used to image blood perfusion in organs, measure blood flow rate in the heart and other organs, and has other applications as well.

Targeting ligands that bind to receptors characteristic of intravascular diseases can be conjugated to microbubbles, enabling the microbubble complex to accumulate selectively in areas of interest, such as diseased or abnormal tissues. This form of molecular imaging, known as targeted contrast-enhanced ultrasound, will only generate a strong ultrasound signal if targeted microbubbles bind in the area of interest. Targeted contrast-enhanced ultrasound can potentially have many applications in both medical diagnostics and medical therapeutics. However, the targeted technique has not yet been approved for clinical use; it is currently under preclinical research and development.

Contents

Bubble echocardiogram

An echocardiogram is an study of the heart using ultrasound. A bubble echocardiogram is an extension of this that uses simple air bubbles as a contrast medium during this study and often has to be requested specifically. Although colour Doppler can be used to detect abnormal flows between the chambers of the heart (e.g. patent foramen ovale) it has a limited sensitivity. When specifically looking for a defect such as this small air bubbles can be used as a contrast medium and injected intravenously, where they travel to the right side of the heart. The test would be positive for an abnormal communication if the bubbles are seen passing into the left side of the heart. (Normally they would exit the heart through the pulmonary artery and be stopped by the lungs.) This form of bubble contrast medium is generated on an ad-hoc basis by the testing clinician by agitating normal saline (e.g. by rapidly and repeatedly transferring the saline between two connected syringes) immediately prior to injection.

Microbubble contrast agents

General features

There are a variety of microbubbles contrast agents. Microbubbles differ in their shell makeup, gas core makeup, and whether or not they are targeted.

  • Microbubble shell: selection of shell material determines how easily the microbubble is taken up by the immune system. A more hydrophilic material tends to be taken up more easily, which reduces the microbubble residence time in the circulation. This reduces the time available for contrast imaging. The shell material also affects microbubble mechanical elasticity. The more elastic the material, the more acoustic energy it can withstand before bursting.[1] Currently, microbubble shells are composed of albumin, galactose, lipid, or polymers.[2]
  • Microbubble gas core: The gas core is the most important part of the ultrasound contrast microbubble because it determines the echogenicity. When gas bubbles are caught in an ultrasonic frequency field, they compress, oscillate, and reflect a characteristic echo- this generates the strong and unique sonogram in contrast-enhanced ultrasound. Gas cores can be composed of air, or heavy gases like perfluorocarbon, or nitrogen.[2] Heavy gases are less water-soluble so they are less likely to leak out from the microbubble to impair echogenicity.[1] Therefore, microbubbles with heavy gas cores are likely to last longer in circulation.

Regardless of the shell or gas core composition, microbubble size is fairly uniform. They lie within in a range of 1-4 micrometres in diameter. That makes them smaller than red blood cells, which allows them to flow easily through the circulation as well as the microcirculation.

Specific agents

  • Optison,a Food and Drug Administration (FDA)-approved microbubble made by GE Healthcare, has an albumin shell and octafluoropropane gas core. The second FDA-approved microbubble, Levovist, made by Schering, has a lipid/galactose shell and an air core.[2]
  • Perflexane lipid microspheres (trade name Imagent or previously Imavist) is an injectable suspension developed by Alliance Pharmaceutical approved by the FDA (in June 2002) for improving visualization of the left ventricular chamber of the heart, the delineation of the endocardial borders in patients with suboptimal echocardiograms. Beside its use to assess cardiac function and perfusion it is also used as an enhancer of the images of prostate, liver, kidney and other organs.[3]
  • Perflutren lipid microspheres (trade name Definity) are composed of octafluoropropane encapsulated in an outer lipid shell.[4]

Targeted microbubbles

Targeted microbubbles are under preclinical development. They retain the same general features as untargeted microbubbles, but they are outfitted with ligands that bind specific receptors expressed by cell types of interest, such as inflamed cells or cancer cells. Current microbubbles in development are composed of a lipid monolayer shell with a perflurocarbon gas core. The lipid shell is also covered with a polyethylene glycol (PEG) layer. PEG prevents microbubble aggregation and makes the microbubble more non-reactive. It temporarily “hides” the microbubble from the immune system uptake, increasing the amount of circulation time, and hence, imaging time.[5] In addition to the PEG layer, the shell is modified with molecules that allow for the attachment of ligands that bind certain receptors. These ligands are attached to the microbubbles using carbodiimide, maleimide, or biotin-streptavidin coupling.[5] Biotin-streptavidin is the most popular coupling strategy because biotin’s affinity for streptavidin is very strong and it is easy to label the ligands with biotin. Currently, these ligands are monoclonal antibodies produced from animal cell cultures that bind specifically to receptors and molecules expressed by the target cell type. Since the antibodies are not humanized, they will elicit an immune response when used in human therapy. Humanizing antibodies is an expensive and time-intensive process, so it would be ideal to find an alternative source of ligands, such as synthetically manufactured targeting peptides that perform the same function, but without the immune issues.

How it works

There are two forms of contrast-enhanced ultrasound, untargeted (used in the clinic today) and targeted (under preclinical development). The two methods slightly differ from each other.

Untargeted CEUS

Untargeted microbubbles, such as the aforementioned Optison or Levovist, are injected intravenously into the systemic circulation in a small bolus. The microbubbles will remain in the systemic circulation for a certain period of time. During that time, ultrasound waves are directed on the area of interest. When microbubbles in the blood flow past the imaging window, the microbubbles’ compressible gas cores oscillate in response to the high frequency sonic energy field, as described in the ultrasound article. The microbubbles reflect a unique echo that stands in stark contrast to the surrounding tissue due to the orders of magnitude mismatch between microbubble and tissue echogenicity. The ultrasound system converts the strong echogenicity into a contrast-enhanced image of the area of interest. In this way, the bloodstream’s echo is enhanced, thus allowing the clinician to distinguish blood from surrounding tissues.

Targeted CEUS

Targeted contrast-enhanced ultrasound works in a similar fashion, with a few alterations. Microbubbles targeted with ligands that bind certain molecular markers that are expressed by the area of imaging interest are still injected systemically in a small bolus. Microbubbles theoretically travel through the circulatory system, eventually finding their respective targets and binding specifically. Ultrasound waves can then be directed on the area of interest. If a sufficient number of microbubbles have bound in the area, their compressible gas cores oscillate in response to the high frequency sonic energy field, as described in the ultrasound article. The targeted microbubbles also reflect a unique echo that stands in stark contrast to the surrounding tissue due to the orders of magnitude mismatch between microbubble and tissue echogenicity. The ultrasound system converts the strong echogenicity into a contrast-enhanced image of the area of interest, revealing the location of the bound microbubbles.[6] Detection of bound microbubbles may then show that the area of interest is expressing that particular molecular, which can be indicative of a certain disease state, or identify particular cells in the area of interest.

Applications

Untargeted contrast-enhanced ultrasound is currently applied in echocardiography. Targeted contrast-enhanced ultrasound is being developed for a variety of medical applications.

Untargeted CEUS

Untargeted microbubbles like Optison and Levovist are currently used in echocardiography.

  • Organ Edge Delineation: microbubbles can enhance the contrast at the interface between the tissue and blood. A clearer picture of this interface gives the clinician a better picture of the structure of an organ. Tissue structure is crucial in echocardiograms, where a thinning, thickening, or irregularity in the heart wall indicates a serious heart condition that requires either monitoring or treatment.
  • Blood Volume and Perfusion: contrast-enhanced ultrasound holds the promise for (1) evaluating the degree of blood perfusion in an organ or area of interest and (2) evaluating the blood volume in an organ or area of interest. When used in conjunction with Doppler ultrasound, microbubbles can measure myocardial flow rate to diagnose valve problems. And the relative intensity of the microbubble echoes can also provide a quantitative estimate on blood volume.

Targeted CEUS

  • Inflammation: Contrast agents may be designed to bind to certain proteins that become expressed in inflammatory diseases such as Crohn’s disease, atherosclerosis, and even heart attacks. Cells of interest in such cases are endothelial cells of blood vessels, and leukocytes:
    • The inflamed blood vessels specifically express certain receptors, functioning as cell adhesion molecules, like VCAM-1, ICAM-1, E-selectin. If microbubbles are targeted with ligands that bind these molecules, they can be used in contrast echocardiography to detect the onset of inflammation. Early detection allows the design of better treatments. Attempts have been made to outfit microbubbles with monoclonal antibodies that bind P-selectin, ICAM-1, and VCAM-1,[2] but the adhesion to the molecular target was poor and a large fraction of microbubbles that bound to the target rapidly detached, especially at high shear stresses of physiological relevance.[7]
    • Leukocytes possess high adhesion efficiencies, partly due to a dual-ligand selectin-integrin cell arrest system.[8] One ligand:receptor pair (PSGL-1:selectin) has a fast bond on-rate to slow the leukocyte and allows the second pair (integrin:immunoglobulin superfamily), which has a slower on-rate but slow off-rate to arrest the leukocyte, kinetically enhancing adhesion. Attempts have been made to make contrast agents bind to such ligands, with techniques such as dual-ligand targeting of distinct receptors to polymer microspheres,[9][10] and biomimcry of the leukocyte’s selectin-integrin cell arrest system,[11] having shown an increased adhesion efficiency, but yet not efficient enough to allow clinical use of targeted contrast-enhanced ultrasound for inflammation.
  • Cancer: cancer cells also express a specific set of receptors, mainly receptors that encourage angiogenesis, or the growth of new blood vessels. If microbubbles are targeted with ligands that bind receptors like VEGF, they can non-invasively and specifically identify areas of cancers.
  • Gene Delivery: Vector DNA can be conjugated to the microbubbles. Microbubbles can be targeted with ligands that bind to receptors expressed by the cell type of interest. When the targeted microbubble accumulates at the cell surface with its DNA payload, ultrasound can be used to burst the microbubble. The force associated with the bursting may temporarily permeablize surrounding tissues and allow the DNA to more easily enter the cells.
  • Drug Delivery: drugs can be incorporated into the microbubble’s lipid shell. The microbubble’s large size relative to other drug delivery vehicles like liposomes may allow a greater amount of drug to be delivered per vehicle. By targeted the drug-loaded microbubble with ligands that bind to a specific cell type, microbubble will not only deliver the drug specifically, but can also provide verification that the drug is delivered if the area is imaged using ultrasound.

Advantages

On top of the strengths mentioned in the medical sonography entry, contrast-enhanced ultrasound adds these additional advantages:

  • The body is 73% water, and therefore, acoustically homogeneous. Blood and surrounding tissues have similar echogenicities, so it is also difficult to clearly discern the degree of blood flow, perfusion, or the interface between the tissue and blood using traditional ultrasound.[2]
  • Ultrasound imaging allows real-time evaluation of blood flow.[12]
  • Ultrasonic molecular imaging is safer than molecular imaging modalities such as radionuclide imaging because it does not involve radiation.[12]
  • Alternative molecular imaging modalities, such as MRI, PET, and SPECT are very costly. Ultrasound, on the other hand, is very cost-efficient and widely available.[6]
  • Since microbubbles can generate such strong signals, a lower intravenous dosage is needed, micrograms of microbubbles are needed compared to milligrams for other molecular imaging modalities such as MRI contrast agents.[6]
  • Targeting strategies for microbubbles are versatile and modular. Targeting a new area only entails conjugating a new ligand.

Disadvantages

In addition to the weaknesses mentioned in the medical sonography entry, contrast-enhanced ultrasound suffers from the following disadvantages:

  • Microbubbles don’t last very long in circulation. They have low circulation residence times because they either get taken up by immune system cells or get taken up by the liver or spleen even when they are coated with PEG.[6]
  • Ultrasound produces more heat as the frequency increases, so the ultrasonic frequency must be carefully monitored.
  • Microbubbles burst at low ultrasound frequencies and at high mechanical indices (MI), which is the measure of the acoustic power output of the ultrasound imaging system. Increasing MI increases image quality, but there are tradeoffs with microbubble destruction. Microbubble destruction could cause local microvasculature ruptures and hemolysis.[5]
  • Targeting ligands can be immunogenic, since current targeting ligands used in preclinical experiments are derived from animal culture.[5]
  • Low targeted microbubble adhesion efficiency, which means a small fraction of injected microbubbles bind to the area of interest.[7] This is one of the main reasons that targeted contrast-enhanced ultrasound remains in the preclinical development stages.

See also

References

  1. ^ a b McCulloch, M., C. Gresser, S. Moos, J. Odabashian, S. Jasper, J. Bednarz, P. Burgess, D. Carney, V. Moore, E. Sisk, A. Waggoner, S. Witt, and D. Adams. Ultrasound contrast physics: A series on contrast echocardiography, article 3. J Am Soc Echocardiogr. 13: 959-67.
  2. ^ a b c d e Lindner, J.R. 2004. Microbubbles in medical imaging: current applications and future directions. Nat Rev Drug Discov. 3: 527-32.
  3. ^ "Perflexane: (AF0150, AFO 150, Imagent, Imavist™)". Drugs in R&D, Volume 3, Number 5, 2002 , pp. 306-309(4) (Adis International). http://www.ingentaconnect.com/content/adis/rdd/2002/00000003/00000005/art00004. Retrieved 2010-03-08. 
  4. ^ Rxlist.com > Definity, Last reviewed 6/16/2008
  5. ^ a b c d Klibanov, A.L. 2005. Ligand-carrying gas-filled microbubbles: ultrasound contrast agents for targeted molecular imaging. Bioconjug Chem. 16: 9-17.
  6. ^ a b c d Klibanov, A.L. 1999. Targeted delivery of gas-filled microspheres, contrast agents for ultrasound imaging. Adv Drug Deliv Rev. 37: 139-157.
  7. ^ a b Takalkar, A.M., A.L. Klibanov, J.J. Rychak, J.R. Lindner, and K. Ley. 2004. Binding and detachment dynamics of microbubbles targeted to P-selectin under controlled shear flow. J. Contr. Release. 96: 473-482.
  8. ^ Eniola, A.O., P.J. Willcox, and D.A. Hammer. 2003. Interplay between rolling and firm adhesion elucidated with a cell-free system engineered with two distinct receptor-ligand pairs. Biophys. J. 85: 2720-31.
  9. ^ Eniola, A.O., and D.A. Hammer. 2005. In vitro characterization of leukocyte mimetic for targeting therapeutics to the endothelium using two receptors. Biomaterials. 26: 7136-44.
  10. ^ Weller, G.E., F.S. Villanueva, E.M. Tom, and W.R. Wagner. 2005. Targeted ultrasound contrast agents: In vitro assessment of endothelial dysfunction and multi-targeting to ICAM-1 and sialyl Lewis(x). Biotechnol. Bioeng. 92: 780-8.
  11. ^ Rychak J.J., A.L. Klibanov, W. Yang, B. Li, S. Acton, A. Leppanen, R.D. Cummings, and K. Ley. "Enhanced Microbubble Adhesion to P-selectin with a Physiologically-tuned Targeting Ligand," 10th Ultrasound Contrast Research Symposium in Radiology, San Diego, CA, March 2005.
  12. ^ a b Lindner, J.R., A.L. Klibanov, and K. Ley. Targeting inflammation, In: Biomedical aspects of drug targeting. (Muzykantov, V.R., Torchilin, V.P., eds.) Kluwer, Boston, 2002; pp. 149–172.

External links


Wikimedia Foundation. 2010.

Игры ⚽ Нужен реферат?

Look at other dictionaries:

  • Contrast medium — A medical contrast medium (or contrast agent) is a substance used to enhance the contrast of structures or fluids within the body in medical imaging.[1] It is commonly used to enhance the visibility of blood vessels and the gastrointestinal tract …   Wikipedia

  • Ultrasound — Not to be confused with Supersonic .Ultrasound is cyclic sound pressure with a frequency greater than the upper limit of human hearing. Although this limit varies from person to person, it is approximately 20 kilohertz (20,000 hertz) in healthy,… …   Wikipedia

  • Double-contrast barium enema — A double contrast barium enema is a procedure in which x rays of the colon and rectum are taken after a liquid containing barium is put into the rectum. Barium is a silver white metallic compound that outlines the colon and rectum on an x ray and …   Wikipedia

  • High-intensity focused ultrasound — Contents 1 Theory 2 Aiming 3 How HIFU works 4 Method of use …   Wikipedia

  • High intensity focused ultrasound — HIFU (high intensity focused ultrasound) (sometimes FUS or HIFUS) is a highly precise medical procedure using high intensity focused ultrasound to heat and destroy pathogenic tissue rapidly. It is one modality of therapeutic ultrasound, and… …   Wikipedia

  • Gravis Ultrasound — or GUS is a sound card for the IBM PC compatible system platform, made by Canada based Advanced Gravis Computer Technology Ltd. It was very popular in the demo scene in the 1990s, due to its superior sound quality compared to similarly priced… …   Wikipedia

  • Medical ultrasonography — This article is about using ultrasound to image the human body. For imaging of animals in research, see Preclinical imaging. Sonography redirects here. For the tactile alphabet called sonography , see Night writing. Medical ultrasonography… …   Wikipedia

  • Doppler echocardiography — Intervention Mitral valve MeSH …   Wikipedia

  • Echogenicity — (misspelled sometimes as echogenecity) is the ability to create an echo, i.e. return a signal in ultrasound examinations. Echogenicity could be increased by intravenously administering gas filled microbubble contrast agent to the systemic… …   Wikipedia

  • Echocardiography — An echocardiogram is a sonography of the heart. Also known as a cardiac ultrasound, it uses standard ultrasound techniques to image two dimensional slices of the heart. The latest ultrasound systems now employ 3D real time imaging.In addition to… …   Wikipedia

Share the article and excerpts

Direct link
Do a right-click on the link above
and select “Copy Link”