Electron capture detector

Electron capture detector

The electron capture detector (ECD) was invented in 1957, by Dr. James E. Lovelock . [cite journal
title = A sensitive detector for gas chromatography
author = J. E. Lovelock
journal = Journal of Chromatography A
year = 1958
volume = 1
issue = 1
pages= 35–46
doi = 10.1016/S0021-9673(00)93398-3
] It is a device for use in gas chromatography that can detect tiny amounts of chemical compounds in the atmosphere and elsewhere. The electron capture detector is used in detecting electron-absorbing components in the output stream of a gas chromatograph.The ECD uses a radioactive Beta particle (electrons) emitter -- a typical source contains a metal foil holding 10 millicuries of Nickel-63. The electrons formed are attracted to a positively charged anode, generating a steady current. As the sample is carried into the detector by a stream of nitrogen or a 5% methane, 95% argon mixture, analyte molecules capture the electrons and reduce the current between the collector anode and a cathode. The analyte concentration is thus proportional to the degree of electron capture, and this detector is particularly sensitive to halogens, organometallic compounds, nitriles, or nitro compounds.

ensitivity

The ECD is 10-1000 times more sensitive than an FID, and one million times more sensitive than a TCD, but has a limited dynamic range and finds its greatest application in analysis of halogenated compounds. The detection limit for electron capture detectors is 5 femtograms per second (fg/s), and the detector commonly exhibits a 10,000-fold linear range. This made it possible to detect halogenated compounds such as pesticides and CFCs, even at levels of only one part per trillion (ppt), thus revolutionizing our understanding of the atmosphere and pollutants.

References

Reviews

*cite journal
title = Selective detectors in gas chromatography
author = M. Krejči and M. Dressler
journal = Chromatographic Reviews
year = 1970
volume = 13
issue = 1
pages= 1–59
doi = 10.1016/0009-5907(70)80005-9

*cite journal
title = Electron capture detection in gas chromatography
author = E. D. Pellizzari
journal = Journal of Chromatography A
year = 1974
volume = 98
issue = 2
pages= 323–361
doi = 10.1016/S0021-9673(00)92077-6

*cite journal
title = The electron capture detector
author = J.E. Lovelock
journal = Journal of Chromatography A
year = 1974
volume = 99
issue = 1
pages= 3–12
doi = 10.1016/S0021-9673(00)90840-9


Wikimedia Foundation. 2010.

Игры ⚽ Нужно решить контрольную?

Look at other dictionaries:

  • electron capture detector — (ECD)    A type of gas chromatographic detector which is sensitive to halogenated hydrocarbons and other molecules capable of easily gaining an electron. Electron capture is not generally used for hydrocarbon detection …   Forensic science glossary

  • electron capture detector — noun An item of laboratory equipment used coupled to a gas chromatograph for the detection and quantification of very minute amounts halogenated organic compounds …   Wiktionary

  • Detector de captura de electrones — Saltar a navegación, búsqueda El detector de captura de electrones es un tipo de detector utilizado en cromatografía de gases. Fue inventado por James Lovelock. Su funcionamiento básico se basa en la emisión de una partícula β (electrón) por… …   Wikipedia Español

  • Detector de captura de electrones — Detector utilizado en cromatografía de gases. Su funcionamiento básico se basa en la emisión de una partícula ß (electrón) por parte de átomos como el 63Ni o tritio adsorbido sobre una placa de platino o titanio. Típicamente, un ECD (electron… …   Enciclopedia Universal

  • Chromatography detector — A chromatography detector is a device used in gas chromatography (GC) or liquid chromatography (LC) to visualize components of the mixture being eluted off the chromatography column. There are two general types of detectors: destructive and non… …   Wikipedia

  • Neutrino detector — The Sudbury Neutrino Observatory, a 12 meter sphere filled with heavy water surrounded by light detectors located 2000 meters below the ground in Sudbury, Ontario, Canada. A neutrino detector is a physics apparatus designed to study neutrinos.… …   Wikipedia

  • Negative electron-transfer dissociation — (NETD) is an ion/ion reaction in which an electron from an anionic specie is transferred to cationic reagent.[1] Following this transfer event, the electron deficient anion undergoes internal rearrangement and fragments. NETD is the ion/ion… …   Wikipedia

  • Reflection high energy electron diffraction — (RHEED) is a technique used to characterize the surface of crystalline materials. RHEED systems gather information only from the surface layer of the sample, which distinguishes RHEED from other materials characterization methods that rely also… …   Wikipedia

  • Scanning electron microscope — These pollen grains taken on an SEM show the characteristic depth of field of SEM micrographs …   Wikipedia

  • chromatography — chromatographer, n. chromatographic /kreuh mat euh graf ik, kroh meuh teuh /, adj. chromatographically, adv. /kroh meuh tog reuh fee/, n. Chem. the separation of mixtures into their constituents by preferential adsorption by a solid, as a column… …   Universalium

Share the article and excerpts

Direct link
Do a right-click on the link above
and select “Copy Link”