Nucleic acid quantitation

Nucleic acid quantitation

In molecular biology, quantitation of nucleic acids is commonly performed to determine the average concentrations of DNA or RNA present in a mixture, as well as their purity. Reactions that use nucleic acids often require particular amounts and purity for optimum performance. There are several methods to establish the concentration of a solution of nucleic acids, including spectrophotometric quantification and UV fluorescence in presence of a DNA dye.

Contents

Spectrophotometric analysis

Nucleic acids absorb ultraviolet light in a specific pattern. In a spectrophotometer, a sample is exposed to ultraviolet light at 260 nm, and a photo-detector measures the light that passes through the sample. The more light absorbed by the sample, the higher the nucleic acid concentration in the sample.

Using the Beer Lambert Law it is possible to relate the amount of light absorbed to the concentration of the absorbing molecule. At a wavelength of 260 nm, the average extinction coefficient for double-stranded DNA is 0.020 (μg/ml)-1 cm-1, for single-stranded DNA it is 0.027 (μg/ml)-1 cm-1, for single-stranded RNA it is 0.025 (μg/ml)-1 cm-1 and for short single-stranded oligonucleotides it is dependent on the length and base composition (estimation 0.030 (μg/ml)-1 cm-1). Thus, an optical density (or "OD") of 1 corresponds to a concentration of 50 μg/ml for double-stranded DNA. This method of calculation is valid for up to an OD of at least 2.[1] A more accurate extinction coefficient may be needed for oligonucleotides; these can be predicted using the nearest-neighbor model.[2]

Conversion Factors for Nucleic Acids

Nucleic Acid Concentration (μg/ml) for 1 A260 unit
dsDNA 50
ssDNA 37
ssRNA 40

Cuvette nucleic acid quantitation

The optical density of samples measured with 10 mm path length standard cuvettes simply has to be multiplied by the conversion factor to determine the concentration of the sample. Example, a dsDNA sample with an OD of 0.9 corresponds to a sample concentration of 45 µg/ml.

Cuvetteless low volume nucleic acid quantitation

Multiple biological applications (e.g. DNA microarray experiment, array CGH, qPCR) imply quantitative and qualitative nucleic acid analysis with minimal sample volumes. Specialized NanoPhotometer[3] offer the possibility to determine sample concentrations cuveteless with submicroliter volumes (starting with 0.3 µl). In addition, due to the reduction of the optical pathlength samples are diluted automatically in comparison to standard cuvette measurements. The respective virtual dilution factors are considered by the software of the instrument. Because the measurements are processed with undiluted samples, the reproducibility of the results is very high. And if desired, samples can be retrieved after the measurement for further processing.

Sample purity

It is common for nucleic acid samples to be contaminated with other molecules (i.e. proteins, organic compounds, other). The ratio of the absorbance at 260 and 280nm (A260/280) is used to assess the purity of nucleic acids. For pure DNA, A260/280 is ~1.8 and for pure RNA A260/280 is ~2.

Protein contamination and the 260:280 ratio

The ratio of absorptions at 260nm vs 280nm is commonly used to assess DNA contamination of protein solutions, since proteins (in particular, the aromatic amino acids) absorb light at 280nm. [1] [4] The reverse, however, is not true — it takes a relatively large amount of protein contamination to significantly affect the 260:280 ratio in a nucleic acid solution.[1] [5]

260:280 ratio has high sensitivity for nucleic acid contamination in protein:

% protein % nucleic acid 260:280 ratio
100 0 0.57
95 5 1.06
90 10 1.32
70 30 1.73

260:280 ratio lacks sensitivity for protein contamination in nucleic acids:

% nucleic acid % protein 260:280 ratio
100 0 2.00
95 5 1.99
90 10 1.98
70 30 1.94

This difference is due to the much higher extinction coefficient nucleic acids have at 260nm and 280nm, compared to that of proteins. Because of this, even for relatively high concentrations of protein, the protein contributes relatively little to the 260 and 280 absorbance. While the protein contamination cannot be reliably assessed with a 260:280 ratio, this also means that it contributes little error to DNA quantity estimation.

Other common contaminants

  • Contamination by phenol, which is commonly used in nucleic acid purification, can significantly throw off quantification estimates. Phenol absorbs with a peak at 270nm and a A260/280 of 1.2. Nucleic acid preparations uncontaminated by phenol should have a A260/280 of around 2.[1] Contamination by phenol can significantly contribute to overestimation of DNA concentration.
  • Absorption at 230nm can be caused by contamination by phenolate ion, thiocyanates, and other organic compounds. For a pure RNA sample, the A260/230 should be around 2, and for a pure DNA sample, the A260/230 should be around 1.8.[6]
  • Absorption at 330nm and higher indicates particulates contaminating the solution, causing scattering of light in the visible range. The value in a pure nucleic acid sample should be zero.[citation needed]
  • Negative values could result if an incorrect solution was used as blank. Alternatively, these values could arise due to fluorescence of a dye in the solution.

Quantification using fluorescent dyes

An alternative way to assess DNA concentration is to use measure the fluorescence intensity of dyes that bind to nucleic acids and selectively fluoresce when bound (e.g. Ethidium bromide). This method is useful for cases where concentration is too low to accurately assess with spectrophotometry and in cases where contaminants absorbing at 260nm make accurate quantitation by that method impossible.

There are two main ways to approach this. "Spotting" involves placing a sample directly onto an agarose gel or plastic wrap. The fluorescent dye is either present in the agarose gel, or is added in appropriate concentrations to the samples on the plastic film. A set of samples with known concentrations are spotted alongside the sample. The concentration of the unknown sample is then estimated by comparison with the fluorescence of these known concentrations. Alternatively, one may run the sample through an agarose or polyacrylamide gel, alongside some samples of known concentration. As with the spot test, concentration is estimated through comparison of fluorescent intensity with the known samples.[1]

If the sample volumes are large enough to use microplates or cuvettes, the dye-loaded samples can also be quantified with a fluorescence photometer.

See also

References

  1. ^ a b c d e Sambrook and Russell (2001). Molecular Cloning: A Laboratory Manual (3rd ed.). Cold Spring Harbor Laboratory Press. ISBN 978-087969577-4. 
  2. ^ Tataurov A.V.; You Y., Owczarzy R. (2008). "Predicting ultraviolet spectrum of single stranded and double stranded deoxyribonucleic acids". Biophys. Chem. 133 (1-3): 66–70. doi:10.1016/j.bpc.2007.12.004. PMID 18201813. 
  3. ^ Kartha, R. Spectrophotometric Quantification of Nano- and Standard-Volume Samples, (2008, October 7), American Biotechnology Laboratory, http://www.iscpubs.com/Media/PublishingTitles/b0608kar.pdf
  4. ^ (Sambrook and Russell cites the original paper: Warburg, O. and Christian W. (1942). "Isolierung und Kristallisation des Gärungsferments Enolase". Biochem. Z. 310: 384–421. )
  5. ^ Glasel J. (1995). "Validity of nucleic acid purities monitored by 260nm/280nm absorbance ratios". BioTechniques 18 (1): 62–63. PMID 7702855. )
  6. ^ "The Analysis of DNA or RNA using Its Wavelengths: 230 nm, 260 nm, 280 nm". Bioteachnology.com. 2010-01-13. http://bioteachnology.com/dna/analysis-dna-rna-wavelengths-230-260-280-nm. Retrieved 2010-03-12. 

External links


Wikimedia Foundation. 2010.

Игры ⚽ Поможем написать реферат

Look at other dictionaries:

  • Quantification of nucleic acids — is commonly used in molecular biology to determine the concentrations of DNA or RNA present in a mixture, as subsequent reactions or protocols using a nucleic acid sample often require particular amounts for optimum performance. There exist… …   Wikipedia

  • Retinoic acid receptor alpha — Retinoic acid receptor, alpha PDB rendering based on 1dkf …   Wikipedia

  • DNA nanotechnology — seeks to make artificial, designed nanostructures out of nucleic acids, such as this DNA tetrahedron.[1] Each edge of the tetrahedron is a 20 base pair DNA double helix, and each vertex is a three arm junction. DNA n …   Wikipedia

  • Plate reader — Microplate Readers (also known as Plate readers) are laboratory instruments designed to detect biological, chemical or physical events of samples in microtiter plates. They are widely used in research, drug discovery, bioassay validation, quality …   Wikipedia

  • Branched DNA assay — In biology, a branched DNA assay is a test for specific nucleic acid chains typically used to detect Retroviruses such as HIVcite journal author = Murphy, D. G.; Gonin, P.; Fauvel, M. year = 1999 title = Reproducibility and Performance of the… …   Wikipedia

  • Magnesium in biology — Magnesium is an essential element in biological systems. Magnesium occurs typically as the Mg2+ ion. It is an essential mineral nutrient for life[1][2][3] and is present in every cell type in every organism. For example, ATP (adenosine… …   Wikipedia

  • Digital polymerase chain reaction — (digital PCR, DigitalPCR, dPCR, or dePCR) is a refinement of conventional polymerase chain reaction methods that can be used to directly quantify and clonally amplify nucleic acids including DNA, cDNA or RNA. The key difference between dPCR and… …   Wikipedia

  • Methods in Enzymology — est une série d ouvrages éditée depuis le milieu des années 1950 par les éditions Academic Press / Elsevier. Cette colossale collection représente la plus importante somme de connaissance dans le domaine de l enzymologie, à l échelle mondiale.… …   Wikipédia en Français

  • Methods in enzymology (collection d'ouvrages) — Methods in Enzymology Methods in enzymology est une série d ouvrages éditée depuis le milieu des années 1950 par les éditions Academic Press / Elsevier. Cette colossale collection représente la plus importante somme de connaissance dans le… …   Wikipédia en Français

  • Analysis — A psychology term for processes used to gain understanding of complex emotional or behavioral issues. * * * 1. The breaking up of a chemical compound or mixture into simpler elements; a process by which the composition of a substance is… …   Medical dictionary

Share the article and excerpts

Direct link
Do a right-click on the link above