Köhler illumination

Köhler illumination

Köhler illumination is a type of specimen illumination used in transmitted- or reflected-light microscopy [ [http://www.olympusmicro.com/primer/anatomy/kohler.html Olympus Microscopy Resource Center - Köhler Illumination] ] . It was designed by August Köhler in 1893, and overcame limitations of previous types of sample illumination techniques (ie: critical illumination). Prior to the advent of Köhler illumination, the filament of the bulb used to illuminate the sample could be visible in the sample plane. This created what is known as a filament image. Various techniques were used to overcome the filament image such as lowering the power of the light source, utilizing an opal bulb, or placing an opal glass diffuser in front of the light source. However, all these techniques, while being semi effective in reducing the filament image, had the effect of reducing the quality and uniformity of light reaching the sample. Reducing the power of the light source and utilizing an opal bulb both caused a narrowing of the spectrum of incident light. For transmitted-light microscopy wide spectrum white light is desired to realize the maximum amount of contrast. Further, adding an opal glass diffuser will cause the light reaching the sample to not be uniform. Uniformity of light is essential to avoid shadows, glare, and incorrect contrast when taking images. Köhler illumination overcomes these limitations by utilizing a collector lens in front of the light source. This lens focuses the light at the condenser diaphragm, and this has the effect of putting the condenser diaphram and the filament image in conjugate planes. Thus, the filament image is no longer conjugate to the image plane, and is no longer visible.

Principles

Condenser Optical Components

* High intensity bulb
* Collector lens
* Field diaphragm
* Condenser diaphragm
* Condenser lens

etting Up Köhler Illumination

1. Focus on the specimen.
2. Close the field diaphragm to its most closed state such that you can see the edges of the diaphragm (may be blurry) in the field of view.
3. Use the condenser focus knobs to bring the edges of the field diaphragm into the best focus possible.
4. Use the condenser-centering screws to center the image of the closed field diaphragm in the field of view.
5. Open the field diaphragm just enough so that its edges are just beyond the field of view.
6. Adjust the condenser diaphragm to introduce the proper amount of contrast into your sample. The amount of contrast added will depend on the sample, however too much contrast can introduce artifacts into your images.
7. Adjust the light intensity as necessary. To adjust light intensity it is best to use a neutral density filter rather than raising or lowering the lightsource power supply. Neutral density filters block all wavelengths of light equally, while changing the power to the light source can reduce the spectrum of incident light creating a yellow/brown light effect.

Used In

* Bright field microscopy
* Phase contrast microscopy
* Dark field microscopy
* Differential interference contrast microscopy
* Polarized light microscopy
* Hoffman modulation contrast microscopy

ee also

* Optical microscopy
* August Köhler

External Sources

[http://microscopy.berkeley.edu/courses/TLM/condenser/kohler.html Köhler illumination setup instructions] from berkeley.edu.

References


Wikimedia Foundation. 2010.

Игры ⚽ Нужен реферат?

Look at other dictionaries:

  • Kohler illumination — The recommended type of optical microscope illumination in which the image of the lamp filament is focused in the lower focal plane of the substage condenser. As opposed to collimated illumination in which the light emitting surface is imaged in… …   Dictionary of molecular biology

  • Köhler — is a German surname and may refer to:*August Köhler, microscopist and inventor of the Köhler illumination *Benjamin Köhler, football player *Ernesto Köhler, flute player and composer *Eva Köhler, First Lady of Germany *Georges J. F. Köhler,… …   Wikipedia

  • Illumination de Köhler — L illumination de Köhler est un système d illumination des échantillons en microscopie optique (transmission et réflexion) introduit en 1893 par August Köhler[1],[2]. Problématique Les précédentes techniques d illumination (illumination critique) …   Wikipédia en Français

  • illumination — 1. Throwing light on the body or a part or into a cavity for diagnostic purposes. 2. Lighting an object under a microscope. [L. il lumino, pp. atus, to light up] axial i. the transmission or reflection o …   Medical dictionary

  • Köhler — Alban, German roentgenologist, 1874–1947. See K. disease. August, German microscopist, 1866–1948. See K. illumination …   Medical dictionary

  • August Köhler — Infobox Scientist name = August Köhler |300px caption = August Köhler, ca. 1930 birth date = birth date|1866|3|4|mf=y birth place = Darmstadt, Germany death date = death date and age|1948|3|12|1866|3|4|mf=y death place = Jena, Germany residence …   Wikipedia

  • Critical illumination — or Nelsonian illumination is a method of specimen illumination used for transmitted and reflected light (trans and epi illuminated) optical microscopy. Critical illumination focuses an image of a light source on to the specimen for bright… …   Wikipedia

  • Maximilian Kohler — Illuminati ist der deutsche Titel eines im März 2003 erschienenen Thrillers von Dan Brown (englischer Originaltitel: Angels and Demons, 2000) mit einer weltweiten Auflage von acht Millionen Exemplaren. Die Handlung rankt sich um die Zerstörung… …   Deutsch Wikipedia

  • Nonimaging optics — (also called anidolic optics)[1][2] is the branch of optics concerned with the optimal transfer of light radiation between a source and a target. Unlike traditional imaging optics, the techniques involved do not attempt to form an image of the… …   Wikipedia

  • Optical microscope — Microscope Uses Small sample observation Notable experiments Discovery of cells Inventor Hans Lippershey Zacharias Jans …   Wikipedia

Share the article and excerpts

Direct link
Do a right-click on the link above
and select “Copy Link”